TS6B 6mm Laser Shutter
The TS6B is available by special order only, and may be subject to minimum order quantities. Please contact us for additional information.
Description
Jump to: Key Features | Specifications | Downloads
The TS6B is available by special order only, and may be subject to minimum order quantities. Please contact us for additional information.
The Uniblitz TS6B is a 6mm bi-stable laser shutter that withholds an extremely low-profile construction. The device’s lack of an external protruding actuator makes it very slim and flexible for system integration. Precision exposure control and laser switching are applications that benefit greatly from the precise, repeatable characteristics of the TS shutter series.
Bi-stable shutter devices, like the TS6B, require no power to hold the blades in either the open or closed state.
Key Features
- 6mm aperture
- Bi-stable operation
- Extremely low-profile form-factor
- RoHS Compliant
- Transfer time on opening: 1.7 milliseconds
- Total opening time: 3.1 milliseconds
- Configured for the VED24 Shutter Driver
Specifications
For additional technical information please see the TS6B Laser Shutter Datasheet.
Coil Resistance | Pulse Voltage to Open | Hold Voltage |
---|---|---|
8 OHMS | +10.7 VDC 1 | N/A |
Weight | Operating Temp. | Max. Opening Bounce | Max. Closing Bounce | Max. Freq. of Operation 2 | Number of Shutter Blades |
---|---|---|---|---|---|
0.17 oz (4.70 g) | -10 – +65 °C | 5% | 5% | 15.0 Hz / 45.0 Hz | 1 |
1 Peak voltage as provided by the VED24 Driver
2 (Continuous/Burst) Continuous frequency rating specified at shutter’s minimum exposure pulse. Burst frequency rating specified for four (4) seconds maximum with one (1) minute minimum between bursts.
Shutter Timing
The question regarding enhancement of shutter speed with the application of user supplied lubricants has been repeatedly asked. It is our experience that lubricating the shutter blades will actually slow the shutter down and eventually render the shutter inoperable. Under no circumstances should any type of lubricant be applied to the shutter blade area.
[/span5][span7]
TS6B (w/ Uniblitz driver and Teflon® coated blades) | Time (msec.) | |
---|---|---|
O – A: | Delay time on opening after current applied | 1.4 |
A – C: | Transfer time on opening | 1.7 |
O – C: | Total opening time | 3.1 |
C – E: | Min. dwell time with min. input pulse | 9.3 |
B – F: | Min. equivalent exp. time | 11.0 |
D – E: | Delay time on closing after current applied | 1.4 |
E – G: | Transfer time on closing | 1.7 |
A – G: | Total window time | 12.7 |
MET: | Min. exposure time | 11.0 |
TEP: | Typical exposure pulse | >11.0 |
[/span7][/columns]
Technical Drawing
Downloads
Documentation | |
---|---|
TS6B Datasheet | Download |
3D Models | |
---|---|
TS6B 3D Outline Model (STEP included in ZIP) | Download |